Pressure instrument
Contact Information
Address: Building 10, Zhonglian Industrial Park, Zhongbei Town, Xiqing District, Tianjin
Telephone:+86-22-59958977
Mailbox:350332639@qq.com
Proline Promass X 500 Coriolis Mass Flow Meter
Classification:
Key words:
- Product Description
-
- Commodity name: Proline Promass X 500 Coriolis Mass Flow Meter
- Commodity ID: Proline Promass X 500
The four-measurement-tube flowmeter with maximum flow capacity, paired with a split-type transmitter offering up to 4-channel input/output capabilities.
Application Areas
The Promass X features four measuring tubes, delivering Premium measurement accuracy (0.05%) and maximum flow capacity, making it ideal for applications on coastal/offshore platforms in the oil and gas industry. Paired with the innovative split-type transmitter, the Promass X 500 maintains exceptional installation flexibility and superior operational safety even under demanding conditions. Additionally, Heartbeat Technology ensures reliable measurements and seamless compliance verification.
- The measurement principle is unaffected by the physical properties of the fluid (e.g., viscosity and density).
- It delivers superior performance in maximum flow measurement applications on coastal/offshore platforms within the oil and gas industry.
Instrument Features
- Nominal diameter: DN 300...400 (12...16")
- Four-rod measurement tube system, low pressure drop
- The entire outer surface is made of 1.4435 (316L) material.
- Split-type instrument, with up to 4 input/output channels
- Backlit display, touch-key control, and WLAN access
- Sensors and transmitters are connected via standard cables.
Advantages
- Easy replacement: Achieve high-flow measurement with premium accuracy from a single installation point.
- Fewer process measurement points: Multi-parameter measurements (flow, density, temperature)
- Save installation space: No straight pipe section length requirements at the front or rear
- View complete process and diagnostic information: Offers multiple user-customized input/output combination methods and Ethernet connectivity.
- Reducing complexity and diversity: User-customizable input/output functionality
- Built-in verification function: Features integrated Heartbeat Technology
Product Specification Parameters
- Maximum measurement error
Mass flow rate (liquid): ±0.10% (standard), 0.05% (optional)
Volumetric flow rate (liquid): ±0.10%
Mass flow rate (gas): ±0.35%
Density (liquid): ±0.0005 g/cm³- Measurement Range
0...4100 t/h (0...4520 tn. sh./h)
- Medium temperature range
–50...+180 °C (–58...+356 °F)
- Maximum Process Pressure
PN 100, Cl. 600
- Liquid-contact component material
Measurement tube: 1.4404 (316/316L)
Connection components: 1.4404 (316/316L)